How to calibrate a load cell scale
Recalibrate a load cell scale after it has been moved, repaired or used for a long time. Calibration uses test weights of known mass so that the weighing indicator reads the true weight.
Before you start
- Have test weights close to the range you normally weigh.
- Power up the weighing indicator and let it warm up until the empty reading is stable.
- Clean the platform and make sure nothing touches or leans on it.
Step 1: zero calibration
With the scale empty, enter calibration mode on the weighing indicator and set zero. If zero keeps drifting, fix vibration, wiring or load cell problems first; calibrating an unstable scale only hides the fault.
Step 2: span calibration
Place the test weight and enter its value. The indicator calculates the conversion factor from the load cell output. A weight that is too light magnifies the error, so use a weight close to your working range.
Step 3: linearity check
Load weights at several points, for example a quarter, half and three quarters of capacity, and record each reading. Every point should be within tolerance.
Step 4: four-corner test
Put the same weight on each corner of the platform in turn. Large differences mean the load cells are not matched or the platform is not evenly supported. Trim the corners in the junction box, then calibrate again.
Calibration is not legal verification
Calibration makes the scale accurate. Scales used for trade, where price depends on weight, also need legal verification under local weights and measures rules.
Our weighing indicators include step-by-step calibration guidance on screen, such as the BDI-9903C check weighing controller.